Aircurve 10 ST

Aircurve 10 ST

Technical Specifications

  • Dimensions BiPAP (H x W x D): 4.6” x 6.5” x 5.7” @ 2.9 lbs
  • Heated Humidifier (H4i) (H x W x D): 5.0” x 6.7” x 8.3” @ 1.6 lbs
  • Symbols which appear on the device: Attention, consult accompanying documents; Drip proof; Type CF equipment; Dangerous voltage; Class II equipment; Start/Stop; Manufacturer.
  • Operating pressure range: 3 to 25 cm H2O
  • Maximum single fault steady state pressure: 40 cm H2O
  • Pressure measurement tolerance: ±0.5 cm H2O ± 4% of the measured reading
  • Flow measurement tolerance: ±0.1 or 20% of reading, whichever is greater
  • VAuto mode: 4 to 25 cm H2O (measured at the mask); Min EPAP: 4 cm H2O; Max IPAP: 25 cm H2O; Pressure Support: 0 to 10 cm H2O S, ST and T modes IPAP: 4 to 25 cm H2O (measured at the mask); EPAP: 3 to 25 cm H2O (measured at the mask);
  • Pressure Support: 0 to 22 cm H2O
  • CPAP mode: 4 to 20 cm H2O (measured at the mask)
  • DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871
  • Sound pressure level: 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1: 2002
  • 28 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
  • Sound power level: 36 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
  • Dimensions (L x W x H): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
  • Weight: 2.9 lbs (1.3 kg)
  • Power supply Input range :100–240 V, 50–60 Hz, 40 VA (typical power consumption), < 100 VA (maximum power consumption)
  • Housing construction: Flame retardant engineering thermoplastic
  • Operating temperature: 41ºF to 95ºF ( 5ºC to 35ºC)
  • Operating humidity: 10–95% non-condensing
  • Storage and transport temperature: -4ºF to 140ºF (-20ºC to 60ºC)
  • Storage and transport humidity: 10–95% non-condensing
  • Operating altitude: Sea level to 8,500’ (2,600 m)
  • Supplemental oxygen: Recommended maximum supplemental oxygen flow: 4 L/min (VAuto mode); 15 L/min (CPAP, S, ST and T modes)
  • Electromagnetic compatibility: Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments
  • Air filter: Two-layered, powder-bonded, polyester non-woven fiber
  • Air tubing: Flexible plastic, 1 x 6’6” (2 m)
  • Air outlet: The 22 mm conical air outlet complies with ISO 5356-1
  • IEC 60601-1 classification Class II (double insulation), Type CF

Note: The manufacturer reserves the right to change these specifications without notice.